Learning Objectives
Introduction to Micro Motion Coriolis
What is a Flowmeter?
Micro Motion Sensors
Micro Motion Transmitters
Flow Operating Principle
Curved Tube Sensor
Tube Vibration
Signal Generation
No Flow - Tube Motion
No Flow - No Coriolis Effect
Flow - Coriolis Effect
Flow - Delta-T
Flow Calibration Factor
Flow Calibration
Volume Flow Measurement
Straight Tube Sensor
Density Operating Principle
Start Over
Flow Operating Principle: Curved Tube - Tube Vibration
< PREV   NEXT >


By vibrating in opposition, the Coriolis flow tubes are balanced and isolated from external vibration or movement of the flowmeter.

In Micro Motion flowmeters that have dual parallel flow tubes, process fluid entering the sensor is split with half of the fluid passing through each flow tube. During operation, a drive coil is energized causing the tubes to oscillate in opposition to one another.