Model 400A Process Hydrocarbon Analyzer
|
 |
- Precision 1% of fullscale
- Digital output display
- Automatic fuel shut-off
- Automatic flame-out indicator
- Seven selectable ranges
- Remote range change control capability
|
|
FEATURES AND APPLICATIONS
Emerson has designed the Model 400A Hydrocarbon Analyzer for high sensitivity, stability and
safety. The Model 400A uses the flame ionization detection method in a field-proven design to provide reliable
performance in analysis in high purity gases and trace levels of hydrocarbon contaminants in ambient air or other gases.
To assure analyzer stability, the Model 400A features an integral temperature control system that maintains the
internal case temperature at 120°F (49°C) to minimize ambient temperature effects upon the flow control system and electronics.
In addition, the flow system is housed in a semi-isolated chamber away from the main system electronics. This allows normal
adjustments to be made without upsetting the temperature. These key features provide optimum stability throughout the entire ambient temperature range of 32° to
100°F (0° to 43 °C).
Designed for convenience and ease of operation, the front panel includes a digital display indicating hydrocarbon content, range
and flame-out status.
The Model 400A is designed to provide a high degree of operational safety. An
automatic fuel shut-off system is included to stop fuel flow to the burner in the event of a flame-out condition. All
tubing upstream of the burner is rigid metallic tubing made with ferrule/nut-type compression fittings.
Should an internal fuel leak develop, an inlet fuel flow restrictor, ventilation holes in the enclosure and an internal circulation fan serve to
dilute and dissipate the hydrogen fuel to a safety factor below 25% of the LEL
of hydrogen - even for a worst case leak. If the sample is flammable, an optional sample flow restrictor and
automatic solenoid valve shut-off kit provides additional protection against an
internal sample leak.
The analyzer is packaged in a single, compact enclosure for panel, relay rack or bench mounting.
A hinged door on the front panel provides convenient access to the controls for zero, span and range change.
Case design permits top, front and rear access for convenient maintenance.
The range switch provides seven selectable ranges of x1, x2.5, x10, x25, x100, x250 and x1,000 signal
attenuation, while also allowing remote range change by means of contact closures. An eighth range position marked
"RMT" enables this remote range change function. In addition to the front panel digital readout, a 0 to 5 VDC (overrange-able to 0 to 10 VDC)
output is available for each range. An optional 4 to 20 mA output is available for connecting recorders, controllers
or data acquisition devices.
The flame ionization detection (FID) method is used in the Model 400A to determine the concentration of total
hydrocarbons present in a gaseous sample. Hydrogen burned in air produces a flame containing a negligible number
of ions; introduction of a hydrocarbon sample into the flame results in a complex ionization creating a large number of ions. A polarizing voltage applied between the burner jet and a
collector produces an electrostatic field. This results in migration of positive ions to the collector and negative
ions to the burner jet establishing an ionization current between the two points. This current is directly
proportional to the hydrocarbon concentration in the flame.
During the operation of the Model 400A, sample is introduced to the burner at a constant flow rate. This rate is determined by the sample
pressure regulator which provides a controlled back pressure at the sample capillary. Bypass flow rates may be adjusted from 350 to 3,000 cc per minute to control system response.
Fuel and air flow rates to the burner are determined by regulating the gas pressures against controlled porosity restrictors.
The regulators are high precision, direct action units compactly designed for precise flow control.
The Model 400A is used to measure total hydrocarbon emissions downstream from solvent recovery or solvent destruction systems to validate efficiency of VOC
reduction system as well as indicate VOC emission concentration.
An additional application for the Model 400A is the determination of traces of hydrocarbon contaminants in high purity gases such as nitrogen, argon, hydrogen
and oxygen. This application is of special importance in the semiconductor industry and is also useful in other
process industries for checking blanketing gases wherever the presence of hydrocarbons is critical.
The Model 400A is ideally suited for monitoring ambient air for hydrocarbons where emissions must be controlled or checked, as in perimeter monitoring.
It is also useful in determining the efficiency of waste gas incinerators and air purifiers of various types.
|
|