X-STREAM General Purpose Compact Gas Analyzer - X2GC


           File Size = 1031 KB
File Size = 11719 KB           

  • Analyzes up to three components
  • New patent-pending IntrinzX™ photometric technology with intrinsic linearity provides high sensitivity, large dynamic ranges and long-term span stability with extended calibration intervals
  • User-friendly operator interface
  • Easy to install, low maintenance and field repairable
  • Supports NDIR, UV, VIS, paramagnetic and electrochemical O2, and thermal conductivity detectors
  • NDIR: Robust microflow detectors
  • NDUV/VIS: Vacuum diode detector for stability and long life
  • O2: Fast response paramagnetic and electrochemical oxygen sensor with long-term stability
  • TC: Quartz-coated stainless steel thermal conductivity cells
  • Solvent-resistant, corrosion-resistant, intrinsically safe and "infallible containment" solutions are available
  • Analog and relay outputs, digital inputs and Ethernet or serial interface with Modbus communication
  • Autocalibration via internal or external valves
  • Optional barometric pressure compensation and flow sensor

DESCRIPTION
The X-STREAM® series of gas analyzers features multi-channel analysis utilizing infrared, ultraviolet and visible (NDIR/UV/VIS) photometry, paramagnetic and electrochemical oxygen, and thermal conductivity sensor technologies.

X-STREAM analyzers can measure up to 3 components and the measuring principles may be ordered in various combinations.

The instrument has an alphanumeric LCD which is operated by 6 keys. Clear text messages (available in 5 languages) and front panel LEDs provide information about the measurement and analyzer status. The analyzer is supplied as a tabletop version, but an accessory kit is available to modify the analyzer to a rack mount version.

X-STREAM compact analyzers, equipped with an external power supply, offer analog outputs, 4 status signal relay outputs (according to NAMUR NE 107) and Modbus communication over an Ethernet or serial interface. Digital inputs and relay outputs are optionally available. The analyzers can be ordered with either screw terminals or connectors for electrical connections.

Standard 19” general purpose, IP66/NEMA 4X wall mountable field housing (with optional pressurization systems) and flameproof variations are also available; see product data sheets 103-912GP.A01, 103-912F.A01 and 103-912FD.A01 for detailed information.

STANDARD SAMPLE GAS COMPONENTS AND MEASURING RANGES
 
 Gas Component Symbol Lowest Range   Highest Range 
AmmoniaNH30 - 300 ppm 0 - 100% 
ArgonAr 0 - 50% 0 - 100% 
Carbon DioxideCO20 - 10 ppm 0 - 100% 
Carbon MonoxideCO0 - 20 ppm 0 - 100% 
EthaneC2H60 - 1,000 ppm 0 - 100% 
EthyleneC2H40 - 400 ppm 0 - 100% 
HeliumHe0 - 10% 0 - 100% 
HexaneC6H140 - 300 ppm 0 - 10% 
HydrogenH20 - 2% 0 - 100% 
MethaneCH40 - 300 ppm 0 - 100% 
   
 Gas Component Symbol Lowest Range   Highest Range 
n-ButaneC4H100 - 800 ppm 0 - 100% 
Nitrogen DioxideNO20 - 250 ppm 0 - 1% 
Nitogen MonoxideNO0 - 250 ppm 0 - 100% 
Nitrous OxideN2O0 - 200 ppm 0 - 100% 
Oxygen (Electrochemical)O20 - 5% 0 - 25% 
Oxygen (Paramagnetic)O20 - 1% 0 - 100% 
PropaneC3H80 - 1,000 ppm 0 - 100% 
PropyleneC3H60 - 10% 0 - 100% 
Sulfur DioxideSO20 - 130 ppm 0 - 100% 
Sulfur HexafluorideSF60 - 20 ppm 0 - 2% 
Water VaporH2O0 - 1,000 ppm 0 - 3% 

APPLICATIONS
  • Trace monitoring for gas purity and air separation measurements
  • Monitoring fermentation gases in biotechnology
  • Quality control of natural gas production and distribution
  • Exhaust gas measurement for burner efficiency control
  • Flue gas analysis of boilers, power plants and incinerators
  • Internal combustion engine emissions
  • Monitoring biogas for landfill applications