Model 6081 Smart Wireless Transmitter for pH & ORP


File Size = 131 KB
  • HIGH ACCURACY AND RELIABILITY for monitoring applications
  • SELF-ORGANIZING NETWORK for high data reliability and network stability
  • INDUSTRY LEADING WIRELESS SECURITY
  • COMPATIBLE WITH WIRELESS HART™ Model 1420 GATEWAY and Emerson Process Management wireless networks
  • EASY TO READ two-line display with easy to use menus
  • HART® 7.0 Digital Communications
  • CONTINUOUS DIAGNOSTICS monitor sensor performance
FEATURES AND APPLICATIONS
The Model 6081-P transmitter is ideal for monitoring applications, especially in hard-to-reach or cost prohibitive locations. Model 6081-P measures pH and ORP and is compatible with most Rosemount Analytical sensors. The transmitter has a rugged, cast aluminum weatherproof and corrosion-resistant enclosure (NEMA 4X and IP65). The transmitter includes a two-line 16-character display with simple and intuitive menu screens. Plain language prompts in six local languages guide the user through the programming and calibration procedures. Wireless HART™ digital communications allows access to AMS (Asset Management Solutions) for live process variables and useful diagnostics and troubleshooting information.

Installation and start-up of the Model 6081 wireless transmitter is simple. Just power the Model 6081-P and assign it to a wireless network with a Wireless HART™ Model 1420 Gateway. The unit will auto-locate the most efficient path to the host and will begin transmitting measurement data immediately.

The Self-organizing network ensures exceptional data reliability and network stability. All of Emerson Process Management’s wireless devices employ Encryption, Authentication, Verification, Anti-Jamming and Key Management to ensure data transmission and security. Rosemount Analytical devices include SmartPower™ technology to reduce power consumption and extend power module life while delivering highly reliable measurements with rich HART® data and diagnostic information.

GENERAL SPECIFICATIONS
Enclosure: Cast aluminum. NEMA 4X (IP65)
Dimensions: 6.55 x 5.40 x 5.15" (166 x 137 x 131mm)
Conduit Openings: 3/4" FNPT
Ambient Temperature: 32 to 122°F (0 to 50°C)
Storage Temperature: -4 to 158°F (-20 to 70°C)
Relative Humidity: 0 to 95% (non-condensing)
Weight: 7 lbs (3.2 kg)
Shipping Weight: 8 lbs (3.6 kg)
RFI / EMI: EN-61326
Digital Communications: HART® 7.0
         WIRELESS SPECIFICATIONS
Output: Wireless enabled HART® 7.0
Transmit Rate: User selectable
Antenna: PBT/PC integrated omni-directional antenna


FUNCTIONAL SPECIFICATIONS
pH Range: 0 to 14
ORP Range: -1400 to +1400mV
Accuracy:±1 mV @ 25°C ±0.01 pH
Repeatability:±1 mV @ 25°C ±0.01 pH




SmartPower™ is a trademark of Emerson Process Management
Wireless HART™ is a trademark of HART Foundation