AGA10 Speed of Sound
Calculations
Daniel Ultrasonic Flow Meters can save
valuable time with real-time AGA10 SOS calculations. The Mark III
Electronics, using live gas composition, pressure and temperature,
calculates the speed of sound every 5 seconds. The AGA10-calculated and
meter-measured SOS values are then averaged over an hour with the
results reported in an hourly data log. This comparison can help
identify possible process upsets and can validate the integrity of the
flow meter, gas analysis and analog measurements.
The Daniel Customer Ultrasonic
Interface (CUI) software automatically includes the AGA10-calculated
SOS in the inspection reports. This valuable feature prevents manual gas
composition data entry errors.
|