X-STREAM® Field Housing Gas Analyzer - X2F
|
|
- Analyzes up to four components
- New patent-pending IntrinzX™ photometric
technology with intrinsic linearity provides high
sensitivity, large dynamic ranges and long-term
span stability with extended calibration intervals
- Wall mountable NEMA 4X/IP66 painted stainless steel housing
- Type approved pressurization systems and non-incendive solution for hazardous areas
- User-friendly operator interface
- Easy to install, low maintenance and field repairable
- Supports NDIR, UV, VIS, paramagnetic and electrochemical O2, and thermal conductivity
- NDIR: Robust microflow detectors
- NDUV/VIS: Vacuum diode detector for stability and long life
- O2: Fast response paramagnetic and electrochemical oxygen sensor with long-term stability
- TC: Quartz-coated stainless steel thermal conductivity cells
- Solvent-resistant, corrosion-resistant, intrinsically safe and "infallible containment" solutions are available
- Extended ambient temperature range: -20 to +50 °C (-4 to +122 °F)
- Tropicalization for high ambient humidity
- Analog and relay outputs, digital inputs and
Ethernet or serial interface with Modbus communication
- Autocalibration via internal or external valves
- Optional barometric pressure compensation, internal sampling pump and flow sensor
|
|
DESCRIPTION
The X-STREAMŽ series of gas analyzers features multi-channel analysis utilizing infrared, ultraviolet and visible (NDIR/UV/VIS) photometry, paramagnetic and electro-chemical oxygen, and thermal conductivity sensor technologies.
X-STREAM analyzers can measure up to 4 components and the measuring principles may be ordered in various combinations. The physical benches can be installed in their own compartment separated from the electronics. A purge is added for handling corrosive and toxic gases to protect the electronics and to provide operator safety.Optional thermostatic control enables measuring lower sample gas concentrations and measurements at higher dew points.
The instrument has an alphanumeric display protected by impact-tested safety glass and is operated manually by 6 keys without the need to open the enclosure. Clear text messages (available in 5 languages) and front panel LEDs provide information about the measurement and analyzer status.
X-STREAM analyzers, equipped with an internal wide range power supply for all world areas, offer analog outputs,
4 status signal relay outputs (according to NAMUR NE 107) and Modbus communication over an Ethernet or serial interface. Digital inputs and relay outputs are optionally
available.
The NEMA 4X/IP66 design allows operation in harsh industrial environments.
An ATEX, CSA-C/US and IECEx approved non-incendive solution allows measuring non-flammable gases in Zone 2/Div. 2 applications without additional safety measures such as pressurization systems.
An ATEX certified pressurization system enables measuring
non-flammable gases in Zone 2 hazardous areas. The pressurization system consists of a microprocessor unit controlling a solenoid gas inlet valve.
The Zone 2 system requires a separate power supply for the controller. The pressurization unit provides two programmable
alarm contacts.
The ATEX certified pressurization system with leakage compensation enables the measurement of non-flammable
gases with air as the purge gas.
Supplied with a CSA-C/US approved Z purge pressurization
system X-STREAM analyzers can be installed in Div. 2 hazardous areas.
Two operation modes are available to allow measuring of non-flammable or flammable gases:
Non-flammable gases can be measured with a leakage compensation mode, using air or inert gas as purge gas.
Flammable gases can be measured with either leakage compensation or continuous dilution mode. Leakage compensation requires inert gas, while continuous dilution enables the use of air for pressurization.
A pressure switch may be used to activate an alarm in case of pressure loss (limit set to 0.15").
Optional flame arrestors are available on request to avoid flame propagation into the process gas line.
X-STREAM gas analyzers are also available as rack mountable/tabletop (1/2 19" or 19") and flameproof variations:
see product data sheets 103-912GC.A01,
103-912GP.A01 and
103-910FD.A01 for detailed information.
STANDARD SAMPLE GAS COMPONENTS AND MEASURING RANGES |
| Gas Component | Symbol | Lowest Range | Highest Range |
| Acetone | CH3COCH3 | 0 - 200 ppm | 0 - 3% |
| Acetylene | C2H2 | 0 - 3% | 0 - 100% |
| Ammonia | NH3 | 0 - 100 ppm | 0 - 100% |
| Argon | Ar | 0 - 50% | 0 - 100% |
| Carbon Dioxide | CO2 | 0 - 10 ppm | 0 - 100% |
| Carbon Monoxide | CO | 0 - 20 ppm | 0 - 100% |
| Chlorine | Cl2 | 0 - 300 ppm | 0 - 100% |
| Ethylene | C2H4 | 0 - 400 ppm | 0 - 100% |
| Helium | He | 0 - 10% | 0 - 100% |
| Hexane | C6H14 | 0 - 300 ppm | 0 - 10% |
| Hydrogen | H2 | 0 - 2% | 0 - 100% |
| Methane | CH4 | 0 - 100 ppm | 0 - 100% |
|
|
| Gas Component | Symbol | Lowest Range | Highest Range |
| n-Butane | C4H10 | 0 - 800 ppm | 0 - 100% |
| Nitrogen Dioxide | NO2 | 0 - 50 ppm | 0 - 10% |
| Nitogen Monoxide | NO | 0 - 150 ppm | 0 - 100% |
| Nitrous Oxide | N2O | 0 - 100 ppm | 0 - 100% |
| Oxygen (Electrochemical) | O2 | 0 - 5% | 0 - 25% |
| Oxygen (Paramagnetic) | O2 | 0 - 1% | 0 - 100% |
| Propane | C3H8 | 0 - 1,000 ppm | 0 - 100% |
| Sulfur Dioxide | SO2 | 0 - 50 ppm | 0 - 100% |
| Sulfur Hexafluoride | SF6 | 0 - 20 ppm | 0 - 2% |
| Toluene | C7H8 | 0 - 300 ppm | 0 - 5% |
| Vinyl Chloride | C2H3Cl | 0 - 2% | 0 - 2% |
| Water Vapor | H2O | 0 - 1% | 0 - 5% |
|
|
APPLICATIONS
Petrochemical process analysis and control
Metallurgical production, hardening and heat treatment processes
Quality control of natural gas production and distribution
Safety measurement for flammable mixtures
Exhaust gas measurement for scrubber and activated carbon filter efficiency control
Flue gas analysis of recovery boilers and process furnaces
|
|