Flame Photometric Detector (FPD) Module
|
 |
For Rosemount Analytical Model 500 and
700 Process Gas Chromatographs
The Flame Photometric Detector (FPD) Module enables the
measurement of trace sulfur compounds when integrated with
Rosemount Analytical Model 700 or Model 500 Process Gas
Chromatographs. The Flame Photometric Detector and associated
electronic boards are installed in temperature controlled flame-proof
enclosures and mounted on a stand complete with flame air and
hydrogen controls. This design eliminates the need for instrument air
and ambient temperature control - greatly reducing installation cost
of the process gas chromatograph. The FPD Module comes fully
integrated into a Rosemount Analytical Model 500 or Model 700 Process
Gas Chromatograph. The FPD Module is ATEX approved for Zone 1
& 2 locations and suitable for installation in Class I, Groups B, C & D,
Division 1 locations.
|
|
 |
FEATURES
- Temperature-controlled Flame Photometric Detector (FPD) for stable sub-ppm level sulfur component measurement
- Sulfur component analysis per ISO 19739 available
- Auto ignition of FPD with automatic hydrogen purge. Makes 10 attempts to light before raising a "flame out" alarm
- Requires no instrument air for purge or oven
- Field-mountable without elaborate shelters for lower installation costs
- Can be combined with other Model 500 or Model 700 process applications
FLAME PHOTOMETRIC DETECTOR (FPD) MODULE SPECIFICATIONS
Power:    115 VAC ±15%, 220 VAC ±15%, 50/60 HZ
Environment:    +10° - 40°C (50° - 104°F)
Dimensions: (without sample system):
    1585mm H x 960 mm W x 645 mm D
    62.5" H x 37.8" W x 25.4" D
    with Model 500 Process Gas Chromatograph
        1650mm H x 960 mm W x 645 mm D
        65" H x 37.8" W x 25.4" D
    with Model 700 Process Gas Chromatograph
        65" H x 37.8" W x 25.4"D
Approximate Weight: (without sample system): 85 kg (187 lbs.)
Safety Certification: ATEX certified, certificate number
SIRA 05ATEX1055X coding EEx d IIC T3 CE marked per
ATEX Directive (94/9/EC)
Carrier Gas: Nitrogen 99.995% minimum
FPD Fuel: Hydrogen, approximately 120 mL/min
FPD Air: Clean, dry, hydrocarbon-free and sulfur-free air, approximately 200mL/min
Requires no instrument air
Detector: Flame Photometric Detector (FPD); Auto ignition with automatic Hydrogen
purge. Ten attempts before raising a "flame out" alarm.
For gas chromatograph specifications please see the Rosemount Analytical Model
500 or Model 700 data sheets
|
TYPICAL MEASUREMENT RANGES FOR COMPONENTS PER ISO 19739
| Component |
Minimum (mg/m3) |
ppmv |
Maximum (mg/m3) |
ppmv |
| H2S |
0.1 |
0.07 |
10 |
6.6 |
| COS |
0.1 |
0.04 |
10 |
3.7 |
| Methyl Mercaptan |
0.1 |
0.05 |
20 |
9.3 |
| Ethyl Mercaptan |
0.1 |
0.04 |
20 |
7.2 |
| t-Butyl Mercaptan |
0.1 |
0.04 |
20 |
7.2 |
| Diethyl Sulfide |
0.1 |
0.02 |
20 |
5.0 |
| Tetrahydrothiophene |
0.1 |
0.03 |
20 |
5.1 |
|