CAT 100 Continuous Gas Analyzer Transmitter
|
 |
- Single or dual channel multi-component analyzer
- CSA-C/US and ATEX-approved for Class I, Zone 1 Group IIB + H2 environments
- IP 55/IP 66 enclosure (designed to meet NEMA 4X for outdoor installation)
- Supports NDIR, paramagnetic O2, electrochemical O2 and thermal conductivity detectors
- Eliminates high-cost, environmentally-controlled shelters
- Simple, fast and easy maintenance: internal analyzer can be removed and replaced within minutes
- Optional sample conditioning plate or internal sampling handling options provide a complete, stand-alone analyzer system
- High-performance, micro-flow NDIR detector, robust solid-state detectors, solvent and corrosion-resistant
paramagnetic oxygen and TC cells are available
- Part of the PlantWeb® field-based architecture and is compatible with DeltaV
- Analog, digital and relay outputs, serial interface and analog and digital inputs enable sophisticated I/O functionality
|
|
PRODUCT DESCRIPTION
Emerson Process Management, the world leader in process gas analyzers, offers the multi-component,
multi-technology CAT 100 Continuous Gas Analyzer Transmitter with flameproof (explosion proof) packaging.
The CAT 100 provides superior performance and has been designed for precise, continuous gas measurements in a
variety of applications.
FIELD-MOUNTABLE HOUSING
The CAT 100 is housed in a Class I, Zone 1 (IIB) + H2 approved transmitter-style enclosure that can be
field mounted in the harshest environmental conditions. The
enclosure can be mounted near the process without requiring ambient
controls or costly and space-consuming shelters. This feature
greatly reduces installation and utility costs while vastly
improving process efficiency.
MULTI-MEASUREMENT CAPABILITY
The CAT 100 can continuously measure
1 or 2 components in a single analyzer using a combination of non-dispersive
infrared (NDIR) spectroscopy, paramagnetic or electrochemical oxygen and thermal
conductivity sensors. This exceptional flexibility has been achieved
by combining the most sophisticated gas detection sensor technology
with microprocessor controlled signal processing.
MEASUREMENTS
The CAT 100 brings together more than 40 years of development and
process application expertise blending the latest sensors, digital
signal processing and software technologies. The CAT 100 can satisfy
the most demanding single or dual analysis requirements. Many gas
components can be measured including: Carbon Monoxide (CO), Oxygen (O2),
Carbon Dioxide (CO2), Water (H2O), Propane (C3H8),
Methane (CH4), Butane (C4H10),
Hydrogen (H2), Hexane (C8H14) and
Ethylene (C2H4).
TOUCHPAD CONTROL
The CAT analyzer's magnetically operated user interface is placed
behind an impact-tested safety glass. It offers the user setup and control
capabilities without needing to open the analyzer housing or shut
off power. All analyzer key functions are accessible including
autocalibration, limit alarms, range selection (O2 and IR with scope function),
diagnostics and much more by simply pointing a magnetic tool at
the appropriate function key.
SAMPLE HANDLING/CONDITIONING OPTION
With the addition of a sample handling and conditioning system
mounted to an integral plate, the CAT 100 becomes a complete stand-alone
in-field, transmitter-style analyzer system. The sample conditioning
plate option is a small and efficiently designed system that is
bolted directly onto the analyzer housing at the factory for a final
one-piece system construction. This one-piece unit can then
easily be mounted to a the process pipe. This feature offers
highly efficient process monitoring and fast installation at low
cost. The sample conditioning system is designed by experienced
Emerson Process Management engineers who will tailor the sample handling
system around your particular process parameters and required
measurements.
Alternately, the CAT 100 design can incorporate internal
sample handling options such as a gas sampling pump, flowmeter, flow
restrictor and solenoid valve block for autocalibration
and remote valve and calibration control. Two pumps, flowmeters and valve blocks
are available as a special solution for parallel tubed benches.
FOUNDATION™ FIELDBUS COMMUNICATIONS OPTIONS
FOUNDATION™ fieldbus offers the latest, most innovative field control
architecture available. A two-wire digital communications protocol
enables individual instruments to communicate together and be
interoperable with other FOUNDATION™ fieldbus capable instruments.
This communications protocol can carry vast amounts of information that
is not accessible by using standard analog or digital data acquisition methods – information
not only about the process, but also about the equipment controlling
it.
|
|